보유기관명 |
춘천바이오산업진흥원 |
보유기관코드 |
|
활용범위 |
|
활용상태 |
|
표준코드 |
|
표준분류명 |
|
시설장비 설명 |
1. Main system 1) Measurement wavelength range: 190.0nm~800.0nm or more 2) Wavelength accuracy: less than +/- 1.0nm 3) Wavelength repeatability: less than +/- 0.1nm 4) Wavelength scanning time:Minimum 0.1sec(full wavelength range) Time Course mode 5) Measurement time: Approx. 1 sec(measurement start display depends on PC 6) Resolution : less than 3nm 7) Detector: photodiode array 8) Element resolution: 1.5nm or less 9) Stray light : Less than 0.1%T at 340nm Less than 0.15%T at 220nm 10) Photometric mode: Single beam multi channel mode 11) Photometric range: Absorbance : -0.500~2.000Abs Transmirrance : 0.0~300.0% or more 12) Photometric accuracy :+/- 0.005Abs(at 1Abs) or less 13) Photometric repeatability :+/- 0.002Abs(at 1Abs) or less 14) Baseline stability: Within 0.001 Abs/h (in spectral measurement at 500nm) : Within 0.003 Abs/30min (in time course measurement at 500nm) 15) Baseline flatness: +/-0.001Abs or less 16) Noise level: 0.002Abs or less 17) Light source: halogen lamp deuterium lamp 18) Monochromator: Uses aberration correcting concave pressed holographic grating 19) Sample chamber : Open type(coverless)시료의 흡광도 측정 |
장비이미지코드 |
http://nfec.ntis.go.kr/storage/images/equip/photo/itep200710/.thumb/1155056414_1.jpg |
장비위치주소 |
강원 춘천시 후평1동 198-53번지 춘천바이오산업진흥원 바이오타운 바이오2동 2층 210 |
NFEC 등록번호 |
NFEC-2007-10-000796 |
예약방법 |
|
카타로그 URL |
|
메뉴얼 URL |
|
원문 URL |
http://www.zeus.go.kr/equip/read?equipId=Z-NTIS-0000164 |