기업조회

본문 바로가기 주메뉴 바로가기

논문 기본정보

DIACON 정도관리물질의 평가

논문 개요

기관명, 저널명, ISSN, ISBN 으로 구성된 논문 개요 표입니다.
기관명 NDSL
저널명 핵의학기술 = The Korean journal of nuclear medicine technology
ISSN 1229-9901,
ISBN

논문저자 및 소속기관 정보

저자, 소속기관, 출판인, 간행물 번호, 발행연도, 초록, 원문UR, 첨부파일 순으로 구성된 논문저자 및 소속기관 정보표입니다
저자(한글) 김병찬,정창기,김상수,박혜미,김지선
저자(영문)
소속기관
소속기관(영문)
출판인
간행물 번호
발행연도 2015-01-01
초록 Purpose Quality control used in laboratory help to sustain precision and accuracy consistently to ensure the Diagnostic and treatment for medical employee. To perform an evaluation in statistical manner, it is recommended to use different quality control material than calibrator from manufacturer, quality control material must include analyzed concentration perimeter, must use more than two different concentration materials and recommended to use materials to indicate the concentration in clinical use. However, currently used quality control material has limit of evaluation due to unconsidered concentration or improperly used to get clinical importance data. Therefore, we compared the value of quality control, DIACON Hormone Level 1, 2, 3 (Scantibodies, USA) and DIACON Tumor Level 1, 2, 3 which applied to clinical concentration range. Materials and Methods DIACON Hormone Level 1, 2, 3 (Scantibodies, USA) and DIACON Tumor Level 1, 2, 3 (Scantibodies, USA) that used materials and methods test to compare the result from five RIA testing institution and one institution that used Elecsys (ROCHE) and Architect (ABOTT) for each level. The reagent RIA kit that used in this experiment is followed by Table 2 and 3. Test has performed by instruction itself and follow by instruction. Results Among DIACON hormone result of twenty four items, level 1 have three items, level 2 have four items and level 3 have five items were out of manufacturer set up range based on the average value. Among DIACON tumor result of thirteen items, level 1 have three items, level 2 have four items and level 3 have five items were out of manufacturer set up range. Other result were inside of manufacturer range. Conclusion For quality control material, it is widely available in market, but it is limited for normal range which is only concerned in improper clinical value and difficult to evaluate important concentration area in terms of clinical analyze. In this experiment, we evaluate Hormone twenty three items and Tumor thirteen items with DIACON Hormone and Tumor (Scantibodies, USA) to resolve this limit and we could observe that it ca be substituted and used.
원문URL http://click.ndsl.kr/servlet/OpenAPIDetailView?keyValue=03553784&target=NART&cn=JAKO201528638840543
첨부파일

추가정보

과학기술표준분류, ICT 기술분류,DDC 분류,주제어 (키워드) 순으로 구성된 추가정보표입니다
과학기술표준분류
ICT 기술분류
DDC 분류
주제어 (키워드) 정도관리,정도관리물질,quality control,DIACON