초록 |
OBJECTIVE A new piezoelectric aptamer biosensor is developed for determination of IgE. The energy converters are 10MHz AT-cut quartz crystals with gold-coated electrodes. The anti-IgE aptamers are immobilized onto the surfaces of crystals gold electrodes by biotin-avidin method. METHODS The standard substance and serum were detected to find the limit of detection and specificity of the biosensor. RESULTS The piezoelectric immunoglobulin aptamer biosensor could complete the detection without nonspecific response. Under the optimized conditions, the experimental results showed that the piezoelectric biosensor had good response to IgE whose frequency shifts were linearly dependent on IgE concentration in different range. The piezoelectric aptamer biosensor had been used to detect IgE in serum, the analytical results given by this method were in satisfactory agreement with those given by chemoluminescence method, its correlation coefficient was 0. 9924. CONCLUSIONS Piezoelectric aptamer biosensor for the determination of IgE is of high sensitivity, high specificity, high analysis speed, unnecessary labeling, simple operation, real-time detection, etc. It is suitable for detection of IgE and should be used for clinical detection. |