저자(한글) |
이상훈,김주리,조삼주,조광환,임천일,김혁주,허현도,신동오,권수일,최진호,관동대학교 의과대학 제일병원 방사선종양학교실,관동대학교 의과대학 제일병원 방사선종양학교실,울산대학교,식품의약품안전평가원,한국식품의약품안전평가원,인하대학교의과대학부속병원,경희대학교,경기대학교 의학물리학과,가천의과대학교 |
초록 |
We have taken surveys about total 72 departments of radiation oncology which is performing the treatment with linear accelerator and brachytherapy unit in Korea. The survey was included the research about the linear accelerator, brachytherapy, Also, we surveyed the various performance (QA period, manpower, time) of quality control for understanding of efficiency. The survey results show that the QA test of daily and weekly are almost same comparing to USA and Europe but the QA performance of monthly and yearly in Korea are 15.5 which is less than USA and Europe recommended QA item number of 17 to 21. The manpower and QA time in Korea also lower than 50% of USA and Europe recommended because the manpower and QA time limitation in Korea. It will be expected that the manual of quality management in each clinic could be appropriately established when combining the present results with previously published AAPM TG-40 and other protocols. |